BAEP: from cochlea to brainstem

These are small waves recorded in the first 10 ms after a brief click. Because they are so small, thousands of stimuli must be averaged; on the other hand, sleep and anaesthesia barely affect them.

Device settings
Stimulus
≈ 100 µs click, per ear
Opposite ear
Masking noise
Intensity
Clearly above hearing threshold (over 60-70 dB in most protocols)
Rate
Usually a non-integer rate around ≈ 10 Hz
Filter
≈ 100 – 3000 Hz
Repetitions
≥ 1000, twice
Analysis time
≈ 10-15 ms
0246810 msIIIIIIIVV
BAEP waves. Those most reliable clinically are I, III and V.
Approximate sources of the waves
WaveApproximate source
IDistal cochlear nerve
IIProximal cochlear nerve
IIILower pons, cochlear nucleus region
IV-VUpper pons – lower midbrain (lateral lemniscus, inferior colliculus)

Interpeak intervals are the key to evaluation, since the middle ear and cochlea affect them less. In adults the I-V interval is generally below about 4.4 ms; the laboratory's value takes precedence.

  • I-III prolongation: between the 8th nerve and lower pons (e.g. vestibular schwannoma).
  • III-V prolongation: upper pons and midbrain.
  • I absent, V normal: may be a peripheral hearing problem; if all waves are lost, check hearing.
  • Reduced V/I amplitude ratio: a weak clue to central involvement.