Upper limb SEP
The median nerve is stimulated electrically at the wrist and the response is followed over the arm, neck and cortex. Intensity is enough to produce a small thumb movement, slightly above motor threshold.
Device settings
- Stimulus
- Median nerve, 0.2 ms square wave, above motor threshold
- Rate
- ≈ 3-5 Hz (non-integer)
- Filter
- ≈ 30 – 1000 Hz (generally up to 3000 Hz)
- Repetitions
- ≥ 500, twice
- Analysis time
- ≈ 40-50 ms
Waves and approximate sources
| Wave | Recording site | Approximate source |
|---|---|---|
| N9 | Erb's point | Brachial plexus |
| N13 | Over C5-C7 | Cervical dorsal horn, dorsal column entry |
| P14 | Scalp, relative to reference | Caudal medial lemniscus |
| N20 | Contralateral scalp | Primary sensory cortex |
Central conduction time (N13-N20) is very valuable, as it excludes peripheral nerve and arm length effects. To see peripheral delay, first look at N9.
Finding → lesion site
| Finding | Site to consider |
|---|---|
| N9 delayed or absent | Peripheral nerve or plexus |
| N9 normal, N13 absent | Root / posterior horn / dorsal column entry |
| N13 normal, N20 absent | Above cervical level, brainstem or thalamocortical pathway |
| N13-N20 prolonged | Slowed central conduction |
Coma prognosis
After cardiac arrest, once the effect of sedation and temperature management has passed, bilateral absence of N20 is among the strong indicators of poor prognosis. The peripheral response (N9, N13) must be intact, and the finding alone is not enough for decisions; multimodal assessment is needed.