VEP: the time of the visual pathway

Pattern-reversal VEP is the standard method. The patient looks at a screen with one eye while a black-and-white checkerboard reverses about twice a second. The main wave of the response recorded occipitally is the positive P100, at about 100 ms in a healthy adult.

Device settings
Active electrode
Oz (O1, O2 for lateral)
Reference
Fz
Stimulus
Monocular, reversing checkerboard
Check size
1° and 0.25°
Reversal rate
≈ 2 reversals/second
Field
At least 15°
Contrast
≥ 80%
Analysis time
At least 250 ms
050100150200250 msN75P100N135
Normal pattern-reversal VEP (positive up): N75, P100 and N135. Amplitude is measured from the N75 peak to the P100 peak.

In interpretation, the difference between the two eyes matters as much as absolute latency. Unilateral P100 prolongation indicates a prechiasmal lesion; the best-known example is optic neuritis, and latency may remain long for years even when amplitude recovers. Bilateral prolongation alone does not localise.

Factors that distort results

Uncorrected refractive error, small pupils, poor fixation, drowsiness and small checks delay or shrink P100. The ISCEV standard requires refractive error to be corrected and visual acuity to be recorded.

Use by stimulus type
MethodWhen
Pattern reversalFirst choice; cooperative patients
Pattern onset/offsetNystagmus or fixation problems; excluding simulation
FlashVery poor vision, infants, uncooperative patients; informs only about pathway integrity