Reading artefacts

The best way to recognise that a signal is not from the brain is to ask whether its distribution is physiological. A finding limited to one electrode or synchronous with body movement is usually an artefact.

Common artefacts
SourceAppearanceClue
Eye blinkLarge, brief frontal deflectionCornea is positive, so Fp1 and Fp2 become positive; bilateral and symmetric
Lateral eye movementOpposite polarity at F7 and F8The side looked towards becomes positive
MuscleFast, irregular high frequencyTemporal and frontal; lessens when the jaw relaxes
ECGSharp wave synchronous with QRSCompare with the ECG channel
PulseSlow wave after QRS at one electrodeSlightly reposition the electrode
Electrode popSudden step in one channelNo field in neighbours
SweatVery slow driftRoom temperature, patient's state
Tongue movementDiffuse slow wavesRepeat while the patient talks
Mains (50 Hz)Regular, fine-toothedCheck impedance and cables
Eye blinkMuscleElectrode pop
Three artefacts: eye blink, muscle and a single-electrode pop.