Interference, artefacts and patient safety

Interference comes from outside; artefacts usually originate from the patient or the system. In both cases the remedy is first to find the source; consider filters only after the source has been addressed.

Reducing 50 Hz interference

  • First lower the impedances and bring them closer together.
  • Keep the cable bundle short, braid it and do not run it parallel to power cables.
  • Check that the ground electrode is sound.
  • Search for the source by switching off the bed, infusion pump, lighting and chargers one by one.
  • In intensive care, power devices from a single outlet where possible; ground loops increase interference.
00.51 sInterferenceNotch filter: spike bluntedSource removed
The same channel with interference suppressed by a notch filter and with the source removed. The filter can also distort the signal.

Stimulus artefact

In conduction studies the stimulus itself creates a large deflection that hides short-latency responses. Placing the ground electrode between stimulus and recording, keeping the skin dry and clean, and rotating the stimulating electrode slightly about the recording axis are usually enough.

Patient safety

  • For electrical stimulation, all devices must share the same earth and be maintained; leakage current matters especially in patients with catheters.
  • Conduction studies are generally considered safe in patients with pacemakers or implanted defibrillators, but stimulate away from the device, keep it brief and follow your institution's protocol.
  • In patients with external (central venous, cardiac) catheters, do not stimulate near the catheter.
  • TMS requires a separate safety screening; see the MEP topic for details.