Factors that mislead
When you see an abnormal result, the first thought must be measurement, not pathology. The following errors are common and create false abnormalities.
Temperature
A cold limb slows sodium channel closure: latency lengthens and velocity falls, but amplitude may increase. Velocity falls by about 1.5-2 m/s for each 1 °C drop. Aim for about ≥ 32 °C in the arm and about ≥ 30 °C in the leg. Warming is more reliable than a correction formula.
Other technical errors
- Inadequate intensity: false low amplitude and conduction block.
- Excessive intensity: co-stimulation of a neighbouring nerve, false normal response.
- Coarse distance measurement: if the elbow is not flexed, ulnar velocity comes out falsely low.
- Electrode position: initial positive dip.
- Distant source: activity of other muscles recorded simultaneously.
Anatomical variants
Martin-Gruber anastomosis: some axons leave the median nerve and cross to the ulnar nerve in the forearm. Median wrist-elbow comparison can show an initial positive dip, and the elbow response can be larger than the wrist one. It is mistaken for conduction block or falsely fast conduction.
Accessory deep peroneal nerve: if the EDB response at the ankle is smaller than with below-knee stimulation, stimulate behind the lateral malleolus and check the response.